Abstract

We present experimental and numerical studies on a method to mitigate screening current-induced field (SCF) for NI REBCO coil. The SCF is the major field error to incorporate a REBCO insert for a high field LTS/HTS magnet. The field-shaking technique is going to be used to mitigate the SCF of 800-MHz REBCO insert magnet (H800) for MIT 1.3-GHz LTS/HTS NMR magnet (1.3 G). The field-shaking using 500-MHz LTS background magnet generates the SCF in H800, due to huge self and mutual inductances of them. In this paper, we tested the effect of the induced current in the NI REBCO coil on the field-shaking technique to mitigate the SCF. The amount of the induced current was decided by the NI REBCO coil status; the open- or closed-loop coil. We performed the three cases of experimental tests and analyzed them. From the test results, we may conclude that we need to limit the ramp rate of L500 during the field-shaking, to minimize the induced current in the HTS insert which consists of the NI REBCO coil.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.