Abstract

The screening current-induced magnetic field in the (Bi,Pb) 2Sr 2Ca 2Cu 3O x (Bi-2223) insert coil proposed for a beyond 1 GHz nuclear magnetic resonance (NMR) spectrometer may generate a long-term field drift, resulting in a loss of field-frequency lock operation and an inability to make high resolution NMR measurements. The measured screening current-induced magnetic field of a Bi-2223 double-pancake coil exhibits a hysteresis effect at 4.2 K that is reproduced by a numerical simulation based on a finite thickness rectangular superconductor bar model. The screening current-induced field at the coil center is of opposite polarity to that generated by the coil current, and thus the apparent field intensity shows a positive drift with time. On the contrary, the field at a coil end is of the same polarity as the coil field, and the apparent field intensity decreases with time. If we wait for ∼1000 h after coil excitation, the field drift rate approaches the field decay rate of the persistent current of 10 −8 h −1, suitable for a long-term NMR measurement in a beyond 1 GHz NMR spectrometer.