Abstract
We tested a simple digital impedance bridge using two nominally equal resistors to form a 1:1 ratio. We focused on resolution and stability of the detectors. Fluctuations of the source voltages were largely removed through postprocessing of the digitized data, and the measurement results were limited by the detector noise. This detector-limited operating condition was first demonstrated using three modified Keysight 3458A multimeters for measurements of the voltage ratios, achieving 0.01 μV/V type A uncertainty in less than 15 min at 1 kHz. In an effort to extend the applicable frequency range and develop a system with off-the-shelf components, we tested a system using three lock-in detectors for measuring small deviations from the perfect AC ratio of unity magnitude, achieving stabilities and resolutions of 0.1 μV/V in a few hours for each point from 1 kHz to 5 kHz.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: Journal of Research of the National Institute of Standards and Technology
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.