Abstract

The NIST Vacuum Double-Crystal Spectrometer (VDCS) has been modernized and is now capable of recording reference-free wavelength-dispersive spectra in the 2 keV to 12 keV x-ray energy range. The VDCS employs crystals in which the lattice spacings are traceable to the definition of the meter through x-ray optical interferometry with a relative uncertainty ﹤10-⁸. VDCS wavelength determination relies upon precision angle difference measurements for which the encoders of the rotation stages have been calibrated using the circle closure method for accurate, absolute angle measurement. The new vacuum-compatible area detector allows quantification of the aberration functions contributing to the observed line shape and in situ alignment of the crystal optics. This latter procedure is augmented with the use of a thin lamella as the first crystal. With these new techniques, x-ray spectra are registered with the VDCS on an absolute energy scale with a relative uncertainty of 10-⁶.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.