Abstract

Commercially available ready to use syringes (Radiometer, Sarstedt, Ciba-Corning and Portex), containing heparinate as an anticoagulant have been tested to evaluate the magnitude of induced preanalytical errors. Tonometered serum pools adjusted to four Ca2+ concentrations were sampled anaerobically. Ca2+ and pH (ICA2 with 3 digits, Radiometer Medical A/S, Denmark; Heparin: anti-Xa factor activity on a chromogenic substrate. Results were expressed as means of 10 measurements and as percentages of the reference values. Sarstedt syringes, (Li-heparinate), yielded a negative bias (-3%). However for 0.5 or 1 mL samples the bias reached -4% to -6%. Radiometer syringes (QS50 and QS90; calcium titrated heparinate) demonstrated biases below -2%. The bias in the Ciba-Corning (Gas-Lyte) syringe was below 2%. Portex (Pulsator) syringes showed biases above +4% even for nominal sampling volumes. All syringes (except Pulsator) released anticoagulant amounts corresponding to the expected values. Radiometer and Ciba-Corning were the only recommendable devices.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.