Abstract

The serum reference material for concentration measurement of ionized calcium in blood has been established by our research group in Japan. The reference method used is the reference standard cell (RSC:IFCC/WGSE, Covington-Umemoto Cell) system. Here we present the results of the evaluation of accuracy using the serum reference material and the hematocrit effect of the nine commercial ionized calcium analyzers being used in routine measurements by Japanese round robin test program. The ionized calcium concentration and pH of two concentration levels of the reference material were analyzed in triplicate measurements according to the standard procedures designated by manufacturers. The effect of hematocrit in ionized calcium concentration measurement was tested using plasma and whole blood with hematocrit values of 35%, 50% and 65%, respectively. The range of bias of ionized calcium values were -0.14 approximately +0.09 mmol/L, -0.12 approximately +0.09 mmol/L for ionized calcium, respectively. The effect of hematocrit in ionized calcium concentration measurement was strongly observed in GEM Premier, CAI-101 and GASTAT-2 analyzers.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.