Abstract

A benchtop high-resolution time-of-flight mass spectrometer (TOF MS) was evaluated for the determination of key organic microcontaminants. The major advantage of the TOF MS proved to be the high mass resolution of about 0.002 Da (10 ppm). Consequently, the detectability of polar pesticides, polynuclear aromatic hydrocarbons and polychlorinated biphenyls is excellent, and detection limits are in the order of 1–4 pg injected mass. Best mass spectral resolution was obtained for medium-scale peaks. It is a disadvantage that the calibration range is rather limited, viz. to about two orders of magnitude. The high mass spectral resolution was especially useful to improve the selectivity and sensitivity when analyzing target compounds in complex samples and to prevent false-positive identifications.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.