Abstract

This work presents a new dual-photoplethysmographic (PPG) system for pulse transit time (PTT) monitoring. An experiment has been set up in order to compare the PTT measurement between carotid and radial arteries from two systems: our physiological multimodal platform (PMP) and the Complior® tonometer. This work explores the comparison between such optical and mechanical modalities. The results show that the PPG device tends to overestimate the PTT (RMSE = 16 ms). Furthermore, both mechanical and optical signals have been superposed and demonstrated that pulse morphologies are quite similar.Clinical Relevance-Carotid-radial pulse wave velocity (PWV) is compared on a small cohort of subjects and significant differences are observed between optical and mechanical-based systems.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.