Abstract
A European comparison of 1 mm hydrophone calibrations in the frequency range 0.5 MHz to 15 MHz is summarized. The project, organized by the Bureau Communautaire de Reference (BCR) of the Commission of the European Communities, involved five participants from different countries measuring the free-field sensitivity of three types of miniature ultrasonic hydrophone. Three different absolute calibration methods were used in the project, which was completed and published in 1991. The uncertainties quoted at the 95% confidence level ranged from 2.3% to 18%, depending on the frequency, method and laboratory. Agreement between laboratories was generally within a few parts in 102 in the range 0.5 MHz to 5 MHz but individual results differed by up to 15% from the grand mean at 10 MHz and 15 MHz.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.