Abstract

The aim of this study is to characterise the neutron flux generated directly behind targets used in medical cyclotrons. The characterisation process aims at determining the feasibility of using the generated neutrons for research purposes in neutron activation analysis. The study was performed by activating gold foils placed directly behind the cyclotron targets. The thermal and epithermal neutron flux were found to be 4.5E+05±8.78E+04neutronscm-2s-1 and 2.13E+06±8.59E+04neutronscm-2s-1, respectively. The flux value is the same order of magnitude listed in the manual produced by the cyclotron manufacturer. The results are encouraging and show high potential for using the cyclotron facility as a thermal neutron source for research purposes. However, it is important radiation protection procedures be followed to ensure the safety of researchers due to the high gamma dose rate measured directly behind the target at 2.46Sv/h using an OSL chip during the beam on time.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.