Abstract

This report presents results from the application of the Monte Carlo N-Particle (MCNP) computer code to the252Cf neutron activation analysis (NAA) Device in the Technical Physics Institute of the Heilongjiang Science Academy of the People's Republic of China. The thermal and epithermal neutron flux at the sample positions and the neutron and photon fluxes on the surfaces of the device were calculated. A comparison between the calculated and experimental thermal and epithermal neutron fluxes at sample positions yield relative errors of less than 10% for the thermal neutron flux.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.