Abstract

More often the embolic materials in the brain create artefacts in the planning CT images that could lead to a dose variation in planned and delivered dose. The aim of the study was to evaluate the dosimetric effect of artefacts generated by the Onyx™ embolization material during Stereotactic Radiosurgery/Radiotherapy (SRS/SRT) planning. An in-house made novel Polymethyl Methacrylate (PMMA) head phantom (specially designed for SRS/SRT plans) was used for this purpose. For the evaluation process, we have created concentric ring structures around the central Onyx materials on both the CT sets (with and without Onyx material). The verification plans were generated using different algorithms namely Analytical Anisotropic Algorithm (AAA), Acuros XB and Monaco based Monte Carlo on both CT sets. Mean integral dose over the region of interest were calculated in both CT sets. The dosimetric results shows, due to the presence of Onyx material, relative variation in mean integral dose to the proximal structure (Ring 1) were −4.02%, −2.98%, and −2.49% for Monte Carlo, Acuros XB, and AAA respectively. Observed variations are attributed to the presence of artefacts due to Onyx material. Artefacts influence the accuracy of dose calculation during the planning. All the calculation algorithms are not equally capable to account such variations. Special cares are to be taken while choosing the calculation algorithms as it impacts the results of treatment outcome.