Abstract
A method is reported for the direct elemental analysis of solutions and liquids by PIXE, utilizing the external proton beam facility at the Van de Graaff accelerator of the Atomic Energy Centre, Dacca. The samples were irradiated in liquid form through a thin Mylar window holding the liquid in a small polyethylene cup. The proton beam was allowed to pass from the vacuum into the air through a 7 μm Be window, then through an air path of 23 mm and the Mylar window of the sample cup, into the solution. Mylar windows up to 1 mg/cm 2 in thickness were used. The X-rays were detected at 90° relative to the beam with a Si(Li) detector. Minimum detection limits and efficiency curves are reported, and the behaviour of some ions during irradiation is discussed. It is shown that under favourable conditions the sensitivity of the method can reach the ppb level.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.