Abstract

Objective: To explore the changes in nickel-titanium (NiTi) archwires after exposure to fluoride-containing acetic acid solutions. Materials and methods: Preformed rectangular NiTi wires were immersed in either an acidulated fluoride solution, a neutral fluoride solution or distilled water (control) for 1.5 hours at 37°C. After immersion, the unloading force values at four deflection points were measured by three-point bending test. Scanning electron microscopy and X-ray diffraction analysis were also performed to characterize the effects of the fluoride exposure on the wire. Results: Unloading force values of NiTi orthodontic wires were significantly decreased after exposure to both fluoride solutions (p < 0.001). Corrosive changes in surface topography were observed for both fluoride solutions. Wires exposed to acidic fluoride appeared as more severely affected. X-ray diffraction analysis showed no change in crystal lattice of NiTi wires in both solutions. Conclusion: The results suggest that using topical fluoride agents with NiTi wire could decrease the functional unloading mechanical properties of the wire and contribute to prolonged orthodontic treatment.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.