Abstract
Background: The aim of this study was to show that virtual surgical planning (VSP) and printed anatomical models support the reconstruction of the center of rotation (COR) and pelvic BRIM during revision hip surgery using a dual-mobility revision cup system in patients with anterior pelvic column damage and soft tissue envelope deficiency. Methods: Patients with anterior pelvic column damage and soft tissue envelope deficiency underwent revision hip arthroplasty. Virtual planning included assessment of bone segmentation, positioning of the cementless revision cup while maintaining the COR, and the assumed inclination and anteversion angles. Results: The diameter of the planned revision cups was 65.5 ± 2.1 mm, and the diameter of the revision cups used was 65.3 ± 2.1 mm. The difference in COR position in the horizontal axis was 7.8 ± 9.3 mm, in the vertical axis was 4.3 ± 5.9 mm, and in the axial plane was 1.6 ± 3.3 mm. The differences in inclination angle and in the anteversion angle were 12.4° and 8.7°, respectively. Conclusions: The use of VSP and 3D models supports the process of performing RHA surgery in patients with damage to the anterior pelvic column and soft tissue envelope deficiency.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.