Abstract
Statistical iterative reconstruction (SIR) using multidetector computed tomography (MDCT) is a promising alternative to standard filtered back projection (FBP), because of lower noise generation while maintaining image quality. Hence, we investigated the feasibility of SIR in predicting MDCT-based bone mineral density (BMD) and vertebral bone strength from finite element (FE) analysis. The BMD and FE-predicted bone strength derived from MDCT images reconstructed using standard FBP (FFBP) and SIR with (FSIR) and without regularization (FSIRB0) were validated against experimental failure loads (Fexp). Statistical iterative reconstruction produced the best quality images with regard to noise, signal-to-noise ratio, and contrast-to-noise ratio. Fexp significantly correlated with FFBP, FSIR, and FSIRB0. FFBP had a significant correlation with FSIRB0 and FSIR. The BMD derived from FBP, SIRB0, and SIR were significantly correlated. Effects of regularization should be further investigated with FE and BMD analysis to allow for an optimal iterative reconstruction algorithm to be implemented in an in vivo scenario.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.