Waveform reproducibility is a critical factor for performing high resolution mass analysis with digitally operated quadrupole mass filters and traps operating in higher stability zones. In this work, Hill equation-based stability calculations were used to define the effect of period jitter on mass analysis in higher stability zones. These calculations correlate well with experimental observations in higher stability zones. Comparison of experiment to theory supplies the basis for defining jitter-based expectations and limits for mass analysis in higher zones.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.