Abstract

We have constructed, and tested as mass filters, linear quadrupoles with added hexapole fields of 4%, 8%, and 12%, with and without added octopole fields. A hexapole field can be added to the field of a linear quadrupole by rotating the two y rods toward an x rod. This also adds an octopole field which can be removed by making the x rods greater in diameter than the y rods. In comparison to conventional quadrupole mass filters these rod sets have severely distorted quadrupole fields, with a mix of both even and odd higher spatial harmonics. They allow evaluating the performance of rod sets with strong geometric and field distortions as mass filters. Conventional mass analysis at the tip of the stability diagram has been compared to mass analysis using islands of stability. The stability islands are produced by applying an auxiliary quadrupole excitation field to the quadrupole. We show that with normal mass analysis at the tip of the stability diagram, the transmission, resolution, and peak shapes are relatively poor in comparison to a conventional rod set. However, the use of islands of stability dramatically improves the resolution and peak shape, and in some cases ion transmission, suggesting that mass analysis with islands of stability may provide a method to overcome a wide range of field imperfections in linear quadrupole mass filters.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.