Abstract

In this report, we present a prototype design of a quadrupole mass filter (QMF) with hyperbolic electrodes, fabricated at the University of Liverpool using digital light processing (DLP), a low-cost and lightweight 3D rapid prototyping (RP) technique. Experimental mass spectra are shown for H(2)(+), D(2)(+), and He(+) ions to provide proof of principle that the DLP mass filter is working as a mass analyzer in the low-mass range (1 to 10 amu). The performance of the DLP QMF has also been investigated for individual spectral peaks. Numerical simulations of the instrument were performed by coupling CPO and Liverpool QMS-2 programs to model both the ion source and mass filter, respectively, and the instrument is shown to perform as predicted by theory. DLP thus allows miniaturization of mass spectrometers at low cost, using hyperbolic (or other) geometries of mass analyzer electrodes that provide optimal ion manipulation and resolution for a given application. The potential of using RP fabrication techniques for developing miniature and microscale mass analyzers is also discussed.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.