Abstract

The aim of this work was to calculate distributions of scatter-to-primary ratios (epsilon(s)/epsilon(p)) and signal-to-noise ratios per pixel (SNRp) in chest images. Such distributions may provide useful information on how physical image quality (contrast, SNR) is distributed over the posterior/anterior (PA) chest image. A Monte Carlo computer program was used for the calculations, including a model of both the patient (voxel phantom) and the imaging system (X-ray tube, anti-scatter grid and image detector). The calculations were performed for three PA thicknesses 20, 24 and 28 cm. For a 24 cm patient, the epsilon(s)/epsilon(p) varies between 0.5 in the lung to 2.5 behind the spine and heart. The corresponding variation of the SNRp is a factor of 3, with the highest values in the lung. Increasing the patient thickness from 20 to 28 cm increases the epsilon(s)/epsilon(p) by a factor of 2.2 behind the spine and heart.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.