Abstract
To assess the diagnostic performance of a new digital chest imaging system based on a selenium detector in the detection of simulated diffuse interstitial pulmonary disease (DIPD). Diffuse interstitial pulmonary disease was simulated by superimposing plastic sheets containing small radio-paque objects (birdseed) on an anthropomorphic chest phantom. the number of sheets was varied from 0 to 4 to simulate the degree of pathology. Eighty conventional radiographs and 80 images with the selenium detector were obtained under comparable conditions. Six observers were asked to indicate the presence of DIPD using a five-level scale of confidence. The total 160 chest images were analyzed using receiver operating characteristic curves. There was no significant difference in detection performance between the conventional and the digital images (95% confidence interval). None of the observers assessed the diagnostic performance of the digital chest images as significantly different from that of the conventional system. The interobserver differences were higher than the differences between digital and conventional chest images.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.