Abstract

The aim of this work was to study the influence of patient thickness, tube voltage and image detector on patient dose, contrast and ideal observer signal-to-noise ratio (SNR(I)), for pathological details positioned at different regions in the image in posterior-anterior (PA) chest radiology. A Monte Carlo computational model was used to compute measures of physical image quality (contrast, SNR(I)) and patient effective dose, E. Two metastasis-like details positioned in the central right lung and right lung near the spine, respectively, were studied. The tube voltage was varied between 100 and 150 kV and the patient thickness between 20 and 28 cm. Both, a computed radiography (CR) system and a direct radiography (DR) system, were investigated. The DR system provides both lower doses and better image quality compared with the CR system. The SNR(I)2/E is approximately 2.9 times higher for the DR system compared with the CR system.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.