Abstract

A 241Am gamma (γ)-ray calibration field that meets the requirements for a γ-ray reference field as specified in the ISO 4037 standard series was established in the Facility of Radiation Standards of the Japan Atomic Energy Agency. The reference air kerma rates were measured using a reference ionization chamber calibrated by the N-80 quality X-ray calibration field of the national metrology standard in Japan and with a correction to account for differences in photon energy due to the calibration field. Conversion coefficients for the 241Am γ-ray calibration field, including those not listed in the ISO 4037 standard series, were calculated based on the measured γ-ray fluence rate spectra.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.