Abstract
Trace quantities of selenium can be determined in the presence of iron, copper and lead using anodic stripping voltammetry, depositing at −0.60 V in 0.1M HClO4 and stripping in the anodic direction. Two separate peaks are observed at −0.25 V and −0.10 V belonging to copper and selenium, respectively. Sometimes one peak may be observed for both copper and selenium. In this case one more stripping (without deposition) must be done to obtain separate peaks. After standard addition, two strippings have to be done also. With this proposed method, 10−7M selenium could be determined as (1.09±0.03) × 10−7M with a 90% confidence interval in blood samples without any separation.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.