Abstract

A method for the determination of selenium in whole blood using a new Zeeman graphite furnace system is presented. A two-part chemical modifier was used in which palladium is reduced to the metallic state in the graphite tube. This reduced palladium modifier gave enhanced thermal stability and sensitivity for selenium compared with the often used nickel modifier. The palladium modifier is injected into the graphite tube prior to adding the blood which has been previously diluted (1 + 4) with a mixture of 0.5%V/V Triton X-100, 0.125%V/V antifoam B and 0.25%m/VL-ascorbic acid, the reducing agent. Optimisation of the analytical parameters is described. Calibration was established using the standard additions method, with additions being made automatically by an autosampler. Precisions were in the range 2.5–8.7% relative standard deviation for levels of selenium in whole blood of 80.5–159 ng ml–1.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.