Abstract

A rapid shrinkage in the minimum feature size of integrated circuits requires analysis of dopants in their shallow source–drain and their extensions with an enhanced depth resolution. Rutherford backscattering spectroscopy (RBS) combining a medium-energy He ion beam with a detector of improved energy resolution should meet the requirement of a depth resolution better than 5 nm at a depth of 10–20 nm in the next 10 years. A toroidal electrostatic analyzer of 4×10 −3 energy resolution has been used to detect the scattered ions of a medium-energy He ion beam. Five keV As + implanted Si or SiO 2 samples were measured. Depth profiling results using the above technique are compared with those of glancing-angle RBS by MeV energy He ions. Limitations in the energy resolution due to various energy-spread contributions have been clarified.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.