Abstract
The ratio error and phase displacement of a current transformer (CT) are measured by using a CT testing system, which consists of a current source, a standard CT, a CT under test, a CT comparator and a CT burden. Methods for on-site calibration of the components in CT testing systems in industry have been developed recently, which do not require any component to be detached from the systems. The method utilizes several travelling transfer standards: standard CTs, non-reactive standard resistors, a calculable wide-ratio error CT and non-reactive shunt resistors. The travelling standard CT is used for absolute evaluation of a current source and a standard CT of industry. The non-reactive standard resistors and the calculable wide-ratio error CT are used to evaluate the linearity of the errors in the CT comparator. The shunt resistors are used to evaluate the CT burden of industry. The on-site calibration method using these travelling transfer standards has been successfully applied to calibration of CT testing systems in industries.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.