Abstract
Voltammetric enzyme immunosensors based on the employment of streptavidin modified screen-printed carbon electrodes (SPCEs) for the detection of rabbit IgG, as a model analyte, were described. Alkaline phosphatase (AP) and 3-indoxyl phosphate (3-IP) were used as the enzymatic label and substrate, respectively. The adsorption of streptavidin was performed by deposition of a drop of a streptavidin solution overnight at 4 °C on the pre-oxidized surface of the SPCEs. The analytical characteristics of these sensors were evaluated using biotin conjugated to AP. The immunosensor devices were based on a specific reaction of rabbit IgG with its biotinylated antibodies, which were immobilised on the modified screen-printed carbon electrodes through the streptavidin:biotin reaction. The immunosensors were used for a direct determination of AP labelled rabbit IgG, and for free rabbit IgG detection using a sequential competitive immunoassay. A calibration curve in the range of 5 × 10 −11 to 1 × 10 −9 M of rabbit IgG was obtained with a estimated detection limit of 5 × 10 −11 M (7.0 ng/ml). These immunosensors were stable for 5 months if they were stored at 4 °C.