Abstract

We report on the design of a data acquisition (DAQ) module for a small animal PET camera developed at our institutes. During the design an important guideline was to develop a system which is built up from strictly identical DAQ modules, and which has no built-in hardware limitation on the maximum number of modules. The developed DAQ module comprises of an LSO scintillator crystal block, a position sensitive PMT, analog signal conditioning circuits, a digitizer, an field programmable gate array (FPGA) for digital signal processing, and a communication module through which the collected data are sent to a cluster of computers for postprocessing and storage. Instead of implementing hardware coincidence detection between the modules, we attach a precise time stamp to each event in our design, and the coincidence is determined by the data collecting computers during postprocessing. The digital CFD algorithm implemented in the FPGA gives a time resolution of 2 ns FWHM for real detector signals

Full Text
Published version (Free)

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call