Abstract

We report on the design of a DAQ module for a small animal PET camera developed at our institutes. During the design an important guideline was to develop a system which is built up from strictly identical DAQ modules, and which has no built-in hardware limitation on the maximum number of modules. The developed DAQ module comprises of an LSO scintillator crystal block, a position sensitive PMT, analog signal conditioning circuits, a digitizer, an FPGA for digital signal processing and a communication module through which the collected data is sent to a cluster of computers for post processing and storage. Instead of implementing hardware coincidence detection between the modules we attach a precise time-stamp to each event in our design, and the coincidence is determined by the data collecting computers during the post processing. The digital CFD algorithm implemented in the FPGA gives a time resolution of 2 to 3 ns FWHM for real detector signals