Abstract

We have developed a new non-conventional electron paramagnetic resonance (EPR) spectrometer, in which no resonant cavity was used. We previously demonstrated a wide frequency range operation of an EPR spectrometer using two loop antennas, one for a microwave transmission and the other for signal detection [1]. In contrast to Ref. [1], the utilization of a microstrip antenna as a transmitter enhanced a capability of wide-band operation. The replacement of conventional capacitors with varactor diodes makes resonance condition easily reproducible without any mechanical action during tuning and matching procedure since the capacitance of the diodes is controlled electronically. The operation of the new EPR spectrometer was tested by measuring a signal of 1,1-diphenil-2-picrylhydrazyl (DPPH) sample in the frequency range of 0.8-2.5 GHz.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.