Abstract

A pulse shape discrimination circuit was designed and used in an experiment measuring double-differential cross sections of (n, charged particle) reaction; to identify p, α and γ. The performance of the circuit was tested. With this circuit, excellent identification of p, α and γ was obtained.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.