Abstract

A methodology has been developed for a dosimetry inter-comparison of intensity modulated radiation therapy (IMRT) delivery in Australasia. The inter-comparison is part of site credentialing for those sites participating in the prostate fractionated irradiation trial (PROFIT) for intermediate-risk prostate patients developed by the Ontario Clinical Oncology Group and coordinated in Australasia by the Trans Tasman Radiation Oncology Group. Features of the dosimetry inter-comparison design included the use of a dedicated pelvic anthropomorphic phantom, the use of a single CT data set of the phantom including contours and the use of radiochromic film as a dosimeter. Action levels for agreement between measured dose and treatment planning system dose have been proposed based on measurement uncertainty and international experience. A trial run of the dosimetry procedure at the reference centre gave results within the predefined action levels.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call