Abstract
Purpose: With intensity modulated radiation therapy (IMRT), the physician can prescribe, design and deliver optimized treatment plans that target the tumor and spare adjacent critical structures. The increased conformity of such plans often comes at the expenses of adding significant complexity to the delivery of the treatment. With volumetrically modulated arc therapy (VMAT), in addition to the modulation of the intensity of the radiation beam, other mechanical parameters such as gantry speed and dose rate are varied during treatment delivery. It is therefore imperative that we develop comprehensive and accurate methods to validate such complex delivery techniques prior to the commencement of the patient’s treatment. Methods : In this study, a Monte Carlo simulation was performed for the high definition multileaf collimator (HD-MLC) of a Varian Novalis TX linac. Our simulation is based on the MCSIM code and provides a comprehensive model of the linac head. After validating the model in reference geometries, treatment plans for different anatomical sites were simulated and compared against the treatment planning system (TPS) dose calculations. All simulations were performed in a cylindrical water phantom as opposed to the patient anatomy, to remove any complexities associated with density effects. Finally, a comparison through gamma analysis of dose plane between the simulation, the TPS and the measurements from the Matrixx array (IBA) was conducted to verify the accuracy of our model against both the measurements and the TPS. Results : Gamma analysis of ten IMRT and ten VMAT cases for different anatomical sites was performed, using a 3%/3 mm passing criterion. The average passing rates were 97.5% and 94.3% for the IMRT and the VMAT plans respectively when comparing the MCSIM and TPS dose calculations. Conclusion : In the present work a Monte Carlo model of a Novalis TX linac which has been tested and benchmarked to produce phase-space files for the treatment head of the linac was used to produce a input phase-space to calculated dose deposition phenomena in different geometries for IMRT and VMAT treatment modalities. The control points defined for the MLC were replaced by blocks with the same characteristics and materials of the linac MLC to speed up the simulation time. With this technique a simulation of a typical IMRT case can be performed with a 10 computer cluster in about 1.02 hours in average. If the number of computer used is increased the computing time can be reduced even more which make our model suitable for clinical use as a second check method to compare the TPS dose calculated. Our results showed that for IMRT and VMAT deliveries with a HD-MLC, there is an average of 95.9% of the points have a gamma index less than 1 with our chosen criterion between our Monte Carlo simulations and the corresponding measurements and TPS calculations in a cylindrical water equivalent phantom. This Monte Carlo code can be used as pre-treatment, independent dose calculation verification for IMRT and VMAT deliveries.
Highlights
Monte Carlo based radiation transport techniques have been beams or as volumetrically modulated used for several years as the most accurate method for dose arc therapy (VMAT)
IMRT cases for different treatment sites were simulated and compared against the dose calculated by the treatment planning system (TPS)
In the RIT software TPS and MCSIM simulation doses were compared against each other and a dose difference of 3% and a distance to agreement (DTA) of 3 mm were selected for the gamma pass-fail criterion
Summary
Monte Carlo based radiation transport techniques have been beams (classic IMRT delivery) or as volumetrically modulated used for several years as the most accurate method for dose arc therapy (VMAT). With the VMAT delivery, in addition to calculation in radiotherapy 1, including plans for intensity the modulation of the radiation beam intensity, the gantry modulated radiation therapy (IMRT) 2-23. With the introduc- speed and the dose rate can vary during treatment. Such tion of IMRT, the treatment planning team can generate very complexity of treatment delivery requires accurate methods conformal plans with high complexity that can be delivered to validate the dose prior to treatment delivery. Either as multi-gantry step-and-shoot or sliding window.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
More From: International Journal of Cancer Therapy and Oncology
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.