Abstract

An HPLC-PDA method was developed and validated for the determination of hydrochlorothiazide in bulk and pharmaceutical formulation. The method was optimized selecting chromatographic conditions of 50 : 50 acetonitrile : water, Inertsil® column (ODS-3 250 mm × 4.6 mm 5 μm), 20 μL injection volume, flow rate of 1 mL/min at ambient temperature (30°C), and 272 nm. Another column of C18Zorbax® (Eclipse Plus, 4.6 × 250 mm, 5 μm) was tested showing no big difference in the method results. The method was validated giving good precision (RSD% < 1), acceptable linearity (R2 ≥ 0.9978), and low LOD and LOQ (0.5 and 1.7 μg/mL, resp.) on both columns. Successful application on pharmaceutical dosage tablet form gave high recovery of 99.93%. The method was compared with official BP and other reported methods. The proposed method is economic, simple, and rapid and hence can be employed for routine analysis in quality control laboratories.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.