Abstract
An on-line zinc preconcentration and determination system implemented with inductively coupled plasma optical emission spectrometry (ICP-OES) was developed. The zinc was precipitated and retained on a minicolumn filled with ethyl vinyl acetate (EVA) at pH 9.0, without using any complexing reagent. The zinc ions were eluted from the minicolumn with 10% (v/v) hydrochloric acid. Experimental conditions including pH and sample loading and eluting variables were evaluated and established. An enrichment factor (EF) of 44 was obtained for Zn 2+ with a preconcentration time of 120 s. Under the optimal conditions, the value of the limit of detection (3 σ) for the preconcentration of 10 mL of sample was 0.08 μg L − 1 . The sampling frequency was about 24 h − 1 . The precision for six replicate determinations (repeatability conditions) at 50 μg L − 1 Zn level was 3.94% relative deviation standard (RSD), calculated from the peak heights obtained. The methodology was successfully applied to the determination of zinc in tap water samples and in a certified VKI reference material QC Metal LL1 DHI (Water & Environment) Denmark.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.