Abstract

An on-line cadmium preconcentration and determination system implemented with inductively coupled plasma optical emission spectrometry (ICP-OES) associated to flow injection (FI) with ultrasonic nebulization system (USN) was studied. The cadmium was retained as the cadmium-2-(5-bromo-2-pyridylazo)-5-diethylaminophenol, Cd-(5-Br-PADAP), complex, at pH 9.5. The cadmium complex was removed from the knotted reactor (KR) with 3.0 mol/L nitric acid. A total enhancement factor of 216 was obtained with respect to ICP-OES using pneumatic nebulization (12 for USN and 18 for KR) with a preconcentration time of 60 s. The value of the detection limit for the preconcentration of 5 mL of sample solution was 5 ng/L. The precision for 10 replicate determinations at the 5 microg/L Cd level was 2.9% relative standard deviation (RSD), calculated from the peak heights obtained. The calibration graph using the preconcentration system for cadmium was linear with a correlation coefficient of 0.9998 at levels near the detection limits up to at least 1,000 microg/L. The method was successfully applied to the determination of cadmium in wine samples.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.