Abstract

Reaction between urea and hypobromite in alkaline solution was found to produce chemiluminescence with a maximum wavelength at 510 nm. A simple chemiluminescence detection method was used for the determination of urea in human urine and natural aqueous samples, which combined this chemiluminescence reaction with a flow injection analysis system. The relative standard deviation for 5 x 10(-7) mol dm-3 urea is 1.9% (n = 6), and the detection limit is 9.0 x 10(-8) mol dm-3 (3Sr). As this chemiluminescence reaction is very fast, a double concentric tube mixer connected directly to the chemiluminescence cell was used to mix urea solution and hypobromite solution. Alkylamines, carboxylic acids and most amino acids do not interfere in the determination. Ammonium ion interferes, but the sensitivity for ammonium ion is only 1% of that for urea. The interference from ammonium ion was removed sufficiently by using an on-line cation-exchange column.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.