This paper introduces a method for determining the characteristic impedance of transmission lines by using the comparison of two calibrations. While the previously reported calibration comparison methods are based on a two-tier calibration, in the proposed method all of the calibration standards are measured using an uncalibrated vector network analyzer (VNA). Previously reported calibration comparison methods compares only calibration techniques implemented using T-parameters. The method proposed in this paper is able to determine the line impedance by comparing calibration techniques implemented using either T- or ABCD-parameters.
Published Version
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.