Abstract

Via multi-dimensional gas chromatography, configured with parallel dual-channel, double detectors, valves switching and back flushing, rapid analysis of the gas compositions consisting of C1–C5 hydrocarbons and permanent gases, such as CO2, H2S, H2, and CO, for direct coal liquefaction has been realized. With four packed chromatographic columns, which are Hayesep-Q pre-column, Hayesep-Q column, molecular sieve 5A column and one PLOT Al2O3 S capillary column, the gas compositions for direct coal liquefaction are analyzed qualitatively and quantitatively by the external standard method. The determination method has such advantages as excellent separation, simple operation, rapid analysis and accurate results.