Abstract

An enantioselective assay of omeprazole in blood plasma using normal-phase liquid chromatography on a Chiralpak AD column and detection by mass spectrometry is described. Omeprazole is extracted by a mixture of dichloromethane and hexane and, after evaporation, redissolution and injection, separated into its enantiomers on the chiral stationary phase. Detection is made by a triple quadrupole mass spectrometer, using deuterated analogues as internal standards. The method enables determination in plasma down to 10 nmol/l (LOQ) and shows excellent consistency suited for pharmacokinetic studies in man.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.