Abstract
Methyl methanesulfonate and ethyl methanesulfonate are potential genotoxic impurities in imatinib mesylate. In this work, a simple, sensitive, reliable, and fast gas chromatography with mass spectrometry method for the simultaneous determination of methyl methanesulfonate and ethyl methanesulfonate was developed and validated. Total analysis time was only 7 min. An n-hexane/water solution was used to dissolve samples, and then extracted-ion-chromatogram mode was used to quantify methyl methanesulfonate and ethyl methanesulfonate. Calibration curves showed good linearity over the studied range for methyl methanesulfonate and ethyl methanesulfonate. The correlation coefficient of fit exceeded 0.999 for each impurity. The LOD and LOQ of methyl methanesulfonate and ethyl methanesulfonate were as low as 0.001 and 0.005 μg/mL, respectively, with RSDs of the peak area within 1.06-1.96%. Method accuracy was within 97.2-99.8% for methyl methanesulfonate and ethyl methanesulfonate. Therefore, this method can be used to quantify methyl methanesulfonate and ethyl methanesulfonate impurities at extremely low levels in imatinib mesylate.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.