Abstract
Methyl methanesulfonate (MMS) and ethyl methanesulfonate (EMS) have been highlighted as potential genotoxic impurities (PGIs). A sensitive LC/MS/MS method is developed and validated for the determination of MMS and EMS impurities in both Lopinavir and Ritonavir Active pharmaceutical ingredient. Method utilizes, Atlantis T3 column with electrospray ionization in multiple reactions monitoring (MRM) mode for quantitation of impurities. The proposed method is specific, linear, accurate and precise. The calibration curves show good linearity over the concentration range of 0.01–0.23 μg/mL for MMS and 0.005–0.23 μg/mL for EMS. The correlation coefficient obtained is >0.99 in each case. Method has very low limit of detection (LOD) and quantification (LOQ). LOD and LOQ of MMS and EMS are as low as ∼0.002 μg/mL and ∼0.01 μg/mL respectively. Method has accuracy within 80–120% for both the analytes. This method is a good quality control tool for quantitation of MMS and EMS impurities at very low levels in Lopinavir and Ritonavir.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.