Abstract

Naloxone, the analyte and the internal standard, sumatriptan, are extracted from plasma using solid-phase extraction columns. Chromatography and detection are performed using isocratic reversed-phase high-performance liquid chromatography (HPLC) with coulometric end-point detection. The standard curve was linear over the range 0–50 ng/ml of naloxone in plasma. The reproducibility, the coefficient of variation (C.V.) of the method over the range of the standard curve was 6.2–11.2%. The recovery averaged 90.4±8.9%. A plasma profile following i.v. administration of naloxone in one normal healthy volunteer is presented.

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.