Abstract

Sumatriptan succinate (the analyte) and naloxone (the internal standard) were extracted from plasma with a solid-phase extraction technique. Chromatography and detection were performed by isocratic reversed-phase high-performance liquid chromatography with coulometric end-point detection. The standard curve was linear over the range 0–100 ng/ml of sumatriptan succinate in plasma. The reproducibility (as defined by the coefficient of variation, C.V.) over the range of the standard curve was 4.9–7.3%. The recovery averaged 83%. The sensitivity was 0.25 ng of sumatriptan on column (allowing a concentration of 0.5 ng/ml to be determined from a 1-ml plasma sample volume). Plasma profiles of the analyte following subcutaneous (s.c.) administration in eight normal male volunteers, are presented.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.