Abstract

Patient dose during diagnostic radiography procedures primarily depends on the entrance surface dose (ESD) and the exposure area. The purpose of this study was to determine ESD for five common diagnostic X-ray examinations using dose area product and X-ray tube output. The ESDs were estimated in a sample of 340 patients using normalised X-ray output from Philips-PrimaryDiagnost digital X-ray machine equipped with digital flat panel scintillator detector. The resultant mean ESDs for chest postero-anterior/lateral, lumbar spine antero-posterior (AP)/lateral and abdomen AP were 0.49, 2.00, 5.74, 12.79 and 4.62mGy, respectively. The mean and third quartile ESD values of this study were significantly higher than the values found in the literature. This can be attributed to the lower tube potential (kVp) values and the higher tube current-time products (mAs). The methodology used here is a better option for low-resources countries for routine dose monitoring.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.