Abstract
A new flow injection analysis (FIA) procedure is proposed for the indirect atomic absorption spectrometric determination of cyanide. The FIA manifold is based on the insertion of the sample into a distilled water carrier, then the sample flows through a solid-phase reactor filled with silver iodide entrapped in polymeric resin beads. The calibration graph is linear over the range 0.2-6.0 mg l-1 of cyanide (correlation coefficient 0.9974), the detection limit is 0.1 mg l-1, the sample throughput is 193 h-1 and the RSD is 0.8%. The method is simple, quick and more selective than other published FIA procedures. The reproducibility obtained by using different solid-phase reactors and solutions is in the range 2.2-3.1% (RSD). The method was applied to the determination of cyanide in commercial samples such as pharmaceutical formulations and industrial electrolytic baths.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.