Abstract
The electroinactiveAlternaria mycotoxins, altenuene and isoaltenuene, are separated by reversed-phase liquid chromatography and derivatized (post-column) with electrochemically generated bromine. Bromine is produced, on-line, by a coulometric cell and the excess of bromine, after reaction with the analyte in the column effluent, is electrochemically detected. The proposed method compares well in term of sensitivity with existing techniques for the determination of altenuene and isoaltenuene in artificially contaminated samples and, in addition, is superior in terms of selectivity. At a signal-to-noise ratio of 3 the detection limit is 8 ng on column. The relationships between generating current, signal and noise are given.
Talk to us
Join us for a 30 min session where you can share your feedback and ask us any queries you have
Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.