A highly sensitive thermospray liquid chromatographic—mass spectrometric method has been developed for the simultaneous determination of FRC-8653 (I), a new calcium antagonist, and its main metabolite (M-4) in plasma. A deuterated analogue of I was added to the plasma as the internal standard. After the purification and concentration of the plasma sample on bonded-phase disposable columns, the extract was injected into the thermospray liquid chromatograph and analysed by selected-ion monitoring mass spectrometry. The calibration curves obtained were linear over the concentration range 0.5–100 ng/ml. The limits of quantification are 0.5 ng/ml for I and 1 ng/ml for M-4 in plasma, which are sufficient to evaluate plasma concentrations after oral administration to rats.

Full Text
Paper version not known

Talk to us

Join us for a 30 min session where you can share your feedback and ask us any queries you have

Schedule a call

Disclaimer: All third-party content on this website/platform is and will remain the property of their respective owners and is provided on "as is" basis without any warranties, express or implied. Use of third-party content does not indicate any affiliation, sponsorship with or endorsement by them. Any references to third-party content is to identify the corresponding services and shall be considered fair use under The CopyrightLaw.