Abstract

Abstract In nuclear facilities, the mandatory airborne contamination surveillance is operated by continuous air monitors (CAMs). It samples the ambient air on a filter and measures the deposited activity. It is designed to trig an alarm whenever the measured activity concentration exceeds a defined threshold. However, in some sites, such as dismantling nuclear sites, a high rate of false alarm is experienced, mainly for artificial alpha. It has been shown that false alarms are directly related to a sudden variation of the aerosol mass sizes distribution, i.e. a wrong detection limit (DL) evaluation. Experiments on the ICARE tests bench have been carried out to compare the CAM’s DL and an estimated DL as a function of the aerosol characteristics, to the measured one. This new estimation shows significant improvements over the previous one and highlights the need to consider aerosol characteristics for a correct airborne contamination measurement.